Buckleys CalChecker Pro’
The BathyCorrometer® Cal-Checker Pro’ is a self-contained, battery-powered instrument used to verify the function of the BathyCorrometer® Pro’ electronics..
Simply insert the BathyCorrometer® Cal-Checker Pro’ into the waterproof bulkhead socket in the BathyCorrometer’s handle to switch the BCM on. The BathyCorrometer® Pro’ will display a reading of .000V, momentarily press the push button switch of the BathyCorrometer® Cal-Checker Pro’ to display a reading of -1.900 volts on the BCM..
Should the BathyCorrometer® Pro’ display a reading outside this range [ -1.900V (+/- 0.002V)]; this indicates the BathyCorrometer® Pro’ electronics may require adjustment, having verified that the BathyCorrometer® Cal-Checker Pro’ is providing the correct output first. To use the BathyCorrometer® Cal-Checker Pro’ with a Mk5 BCM connect the short calibration lead to the 2-pin socket of the Cal-Checker. Insert the BathyCorrometer® Cal-Checker Pro’ into the waterproof bulkhead socket in the BathyCorrometer®’s handle and connect the spring clip to the pointed probe of the BCM, proceed as detailed above to operate the BathyCorrometer® Cal-Checker Pro’..



















Add to Quote